Difference between revisions of "CHX:Hints"
KevinYager (talk | contribs) (→=WAXS Detector Motion) |
KevinYager (talk | contribs) (→Instrument) |
||
Line 41: | Line 41: | ||
*** Positive motion moves sample up; i.e. beam moves down along sample. | *** Positive motion moves sample up; i.e. beam moves down along sample. | ||
*** Negative motion moves sample down; i.e. beam moves up along sample. | *** Negative motion moves sample down; i.e. beam moves up along sample. | ||
− | * Diffractometer also has | + | ** diff_zh : moves sample along beam. |
+ | |||
+ | * Diffractometer also has stages on a "vertical" tower. Because these stages are sitting on a rotation axis (delta-like rotation), their translations (relative to sample) depend on delta. | ||
+ | ** diff_xv : moves sample inboard/outboard. | ||
+ | *** Positive motion moves stage outboard (left, beam POV); i.e. beam moves right along object. | ||
+ | *** Negative motion moves sample inboard (right, beam POV); i.e. beam moves left along object. | ||
+ | ** diff_yv : moves sample up/down. Assuming rotation stage is 180°: | ||
+ | *** Positive motion moves sample down; i.e. beam moves up along sample. | ||
+ | *** Negative motion moves sample up; i.e. beam moves down along sample. | ||
+ | ** diff_zv : moves sample along beam. | ||
+ | |||
===WAXS Detector Motion=== | ===WAXS Detector Motion=== | ||
* delta : goniometer arm arcing up | * delta : goniometer arm arcing up | ||
* gamma : goniometer in-plane rotation | * gamma : goniometer in-plane rotation | ||
+ | |||
+ | ===WAXS Detector=== | ||
+ | * Eiger1M | ||
+ | * To start: | ||
+ | ** Array callbacks: Enable | ||
+ | * To debug: | ||
+ | ** http_proxy= google-chrome | ||
+ | ** http://10.11.1.3/data/ | ||
+ | |||
+ | ===Cameras=== | ||
+ | * camonitor | ||
+ | * XF:11IDB-BI{Cam:08}Stats1:Total:RBV |
Revision as of 12:52, 31 July 2015
This page provides brief notes/hints for using the CHX beamline at NSLS-II.
Contents
General
Status
Contact
- Beamline: x1411
- Floor coordinator: x5046
- Control room: x2555
Controls
ophyd
- Start:
- cd ~
- ./ophyd
- Examples:
- wh_pos() : motor positions
- mov(diff_xh, 1.0) : move motor
- movr(diff_xh, -1.0) : relative motion
- movr([diff_xh, diff_yh], [-1.0,-1.0]) : multiple motions
Control System Studio (CSS)
- Start:
- run-css &
- Use:
- Ctrl+mousewheel : zoom
Replay
- ./replay
- Allows access to log entries and graphs.
Instrument
Sample Motion
- Sample motion uses diffractometer. Sample stage is on "horizontal controls" (diff_xh, diff_yh).
- diff_xh : moves sample inboard/outboard.
- Positive motion moves sample outboard (left, beam POV); i.e. beam moves right along sample.
- Negative motion moves sample inboard (right, beam POV); i.e. beam moves left along sample.
- diff_yh : moves sample up/down.
- Positive motion moves sample up; i.e. beam moves down along sample.
- Negative motion moves sample down; i.e. beam moves up along sample.
- diff_zh : moves sample along beam.
- diff_xh : moves sample inboard/outboard.
- Diffractometer also has stages on a "vertical" tower. Because these stages are sitting on a rotation axis (delta-like rotation), their translations (relative to sample) depend on delta.
- diff_xv : moves sample inboard/outboard.
- Positive motion moves stage outboard (left, beam POV); i.e. beam moves right along object.
- Negative motion moves sample inboard (right, beam POV); i.e. beam moves left along object.
- diff_yv : moves sample up/down. Assuming rotation stage is 180°:
- Positive motion moves sample down; i.e. beam moves up along sample.
- Negative motion moves sample up; i.e. beam moves down along sample.
- diff_zv : moves sample along beam.
- diff_xv : moves sample inboard/outboard.
WAXS Detector Motion
- delta : goniometer arm arcing up
- gamma : goniometer in-plane rotation
WAXS Detector
- Eiger1M
- To start:
- Array callbacks: Enable
- To debug:
- http_proxy= google-chrome
- http://10.11.1.3/data/
Cameras
- camonitor
- XF:11IDB-BI{Cam:08}Stats1:Total:RBV