Line 1: |
Line 1: |
| In transmission-[[SAXS]] ([[TSAXS]]), the x-ray beam hits the sample at normal incidence, and passes directly through without [[refraction]]. TSAXS is normally considered in terms of the one-dimensional [[momentum transfer]] (''q''); however the full 3D form of the ''q''-vector is necessary when considering [[scattering]] from anisotropic materials. The ''q''-vector in fact has three components: | | In transmission-[[SAXS]] ([[TSAXS]]), the x-ray beam hits the sample at normal incidence, and passes directly through without [[refraction]]. TSAXS is normally considered in terms of the one-dimensional [[momentum transfer]] (''q''); however the full 3D form of the ''q''-vector is necessary when considering [[scattering]] from anisotropic materials. The ''q''-vector in fact has three components: |
| :<math> | | :<math> |
− | \mathbf{q} = \begin{bmatrix} q_x & q_y & q_z \end{bmatrix} | + | \mathbf{q} = \begin{bmatrix} q_x \\ q_y \\ q_z \end{bmatrix} |
| </math> | | </math> |
| This vector is always on the surface of the [[Ewald sphere]]. Consider that the [[x-ray]] beam points along +''y'', so that on the [[detector]], the horizontal is ''x'', and the vertical is ''z''. We assume that the x-ray beam hits the flat 2D area detector at 90° at detector (pixel) position <math>\scriptstyle (x,z) </math>. The scattering angles are then: | | This vector is always on the surface of the [[Ewald sphere]]. Consider that the [[x-ray]] beam points along +''y'', so that on the [[detector]], the horizontal is ''x'', and the vertical is ''z''. We assume that the x-ray beam hits the flat 2D area detector at 90° at detector (pixel) position <math>\scriptstyle (x,z) </math>. The scattering angles are then: |
Revision as of 10:57, 30 December 2015
In transmission-SAXS (TSAXS), the x-ray beam hits the sample at normal incidence, and passes directly through without refraction. TSAXS is normally considered in terms of the one-dimensional momentum transfer (q); however the full 3D form of the q-vector is necessary when considering scattering from anisotropic materials. The q-vector in fact has three components:
This vector is always on the surface of the Ewald sphere. Consider that the x-ray beam points along +y, so that on the detector, the horizontal is x, and the vertical is z. We assume that the x-ray beam hits the flat 2D area detector at 90° at detector (pixel) position . The scattering angles are then:
where is the sample-detector distance, is the out-of-plane component (angle w.r.t. to y-axis, rotation about x-axis), and is the in-plane component (rotation about z-axis). The alternate angle, , is the elevation angle in the plane defined by .
Total scattering
The full scattering angle is:
The total momentum transfer is:
Given that:
We can also write:
Where we take for granted that q must be positive.
In-plane only
If (and ), then , , and:
The other component can be thought of in terms of the sides of a right-triangle with angle :
Summarizing:
Out-of-plane only
If , then , , and:
Components
The momentum transfer components are:
Check
As a check of these results, consider:
Where we used:
And, we further note that:
Continuing: