===Energy===
* Can perform experiments from ~2 keV to 24 keV.
+
* Can perform experiments from 2.1 keV to 24 keV.
 
* Energy resolution ([[double crystal monochromator]]): 10<sup>−4</sup>
 
* Energy resolution ([[double crystal monochromator]]): 10<sup>−4</sup>
  
 
===q-range===
 
===q-range===
TBD
+
* q-min: With some effort, at 16 keV the beamline can reach a minimum q of 0.0004 Å<sup>−1</sup> (realspace distance of 1.5 microns).
 +
** In principle, at 2 keV one could reach q = 0.00006 Å<sup>−1</sup> (realspace distance of 9 microns).
 +
* q-max: The WAXS detector can reach 90°, which (for 24 keV) corresponds to 17 Å<sup>−1</sup> (realspace distance of 0.4 Å).
  
 
===Resolution===
 
===Resolution===
Line 24: Line 27:
  
 
===Beam Size===
 
===Beam Size===
* Focusing (H and V) available to ~2 µm
+
* Focusing available to ~2.5 µm (V) and 25 µm (H)
 
 
  
 
==Access==
 
==Access==

Latest revision as of 07:44, 29 August 2019

The Soft Matter Interfaces (SMI) beamline at NSLS-II is a synchrotron undulator-based instrument focused on structural measurements of surfaces and interfaces. The instrument is optimized for x-ray scattering in grazing-incidence mode; i.e. grazing-incidence small-angle x-ray scattering (GISAXS) and grazing-incidence wide-angle x-ray scattering (GIWAXS/[[GIXD]). SMI can also perform transmission-mode scattering measurements (SAXS/WAXS).

Capabilities

CHX can perform a variety of experiments:

Performance

Energy

q-range

  • q-min: With some effort, at 16 keV the beamline can reach a minimum q of 0.0004 Å−1 (realspace distance of 1.5 microns).
    • In principle, at 2 keV one could reach q = 0.00006 Å−1 (realspace distance of 9 microns).
  • q-max: The WAXS detector can reach 90°, which (for 24 keV) corresponds to 17 Å−1 (realspace distance of 0.4 Å).

Resolution

TBD

Flux

At sample position:

  • TBD

Beam Size

  • Focusing available to ~2.5 µm (V) and 25 µm (H)

Access

SMI has taken first light, and is currently undergoing technical commissioning. Access will be available through the NSLS-II proposal system. Interested users should contact beamline staff.